Instructions for Glutamate Decarboxylase (GAD) Detection Kit (Colloidal Gold Method)

Glutamate decarboxylase (GAD) detection kit (colloidal gold method)
Description book

【Product name】 Glutamate decarboxylase (GAD) antibody detection kit (colloidal gold method)
[Packing Specifications] 24 servings / box 48 servings / box [Expected Use] For detecting glutamate decarboxylase antibody (IgM / IgG) in test serum
[Test principle] Using glutamic acid decarboxylase as the antigen solid phase nitrocellulose membrane, the principle of diafiltration indirect method is used to detect serum glutamate decarboxylase antibody.
[Main components]
1. 24 or 48 reaction plates
2. Reagent â…  1 bottle of 0.02mol / L PH 7.4 PBS
3. Reagent Ⅱ 1 bottle of colloidal gold marker [storage conditions and expiration date] The product should be stored at 2 ℃ ~ 8 ℃ conditions, can not be frozen; the expiration date is 24 months.
[Sample requirements]
1. Serum samples cannot be hemolyzed, and should be stored in fresh serum or stored at 2 ℃ ~ 8 ℃ for no more than one week.
2. Hyperlipidemia serum cannot be used.
【Testing method】
1. Drop two drops of reagent â…  into the central hole of the reaction plate and wait for complete infiltration;
2. Drip 100µl of serum into the well of the reaction plate, to be completely infiltrated;
3. Add three drops of reagent â…¡ to the well of the reaction plate until it is completely infiltrated;
4. Infiltrate three drops of reagent â…  into the well of the reaction plate and wait for complete infiltration.
[Interpretation of results]
Positive: red round spots appear on the C-terminal and red round spots on the T-terminal of the reaction plate well, which is positive for glutamate decarboxylase antibody;
Negative: There are red round spots on the C-terminal and no red round spots on the T-terminal of the reaction plate well, which is negative for glutamate decarboxylase antibody.
Failure: No red round spots appear on the C-end of the reaction plate well, or no red round spots appear on the C-end and T-end.
[Limitations of inspection methods]
1. This reagent test is only used to detect glutamate decarboxylase antibody instead of directly detecting glutamate decarboxylase, so a positive result cannot confirm the diagnosis of diabetic disease. The diagnosis of the patient's condition should be combined with a comprehensive analysis of the patient's clinical signs and symptoms and test results.
2. It is possible that serum samples with low antibody content cannot be detected. Some diabetics do not produce antibodies or produce a small amount of antibodies. At this time, a negative result may be displayed.
3. When the test result is suspicious, multiple other items should be tested to confirm the diagnosis.
【Product Performance Index】
1. In-batch precision: both the positive coincidence rate and the negative coincidence rate should be ≥95%, and the color depth of the reaction spots should be close.
2. Precision between batches: Both the positive and negative coincidence rates should be ≥95%.
【Precautions】
1. This product has not yet obtained a product registration certificate number, which is for research only and does not require clinical diagnosis.
2. Once the test is started, it should be carried out continuously according to the operation steps until the end.
3. When the kit is removed from the refrigerator, the reagents should be returned to room temperature.